Keithley 2700 MultiMeter/Switch System

class pymeasure.instruments.keithley.Keithley2700(adapter, name='Keithley 2700 MultiMeter/Switch System', **kwargs)

Bases: KeithleyBuffer, SCPIMixin, Instrument

Represents the Keithley 2700 Multimeter/Switch System and provides a high-level interface for interacting with the instrument.

keithley = Keithley2700("GPIB::1")
beep(frequency, duration)

Sounds a system beep.

Parameters
  • frequency – A frequency in Hz between 65 Hz and 2 MHz

  • duration – A time in seconds between 0 and 7.9 seconds

property buffer_data

Get a numpy array of values from the buffer.

property buffer_points

Control the number of buffer points. This does not represent actual points in the buffer, but the configuration value instead.

channels_from_rows_columns(rows, columns, slot=None)

Determine the channel numbers between column(s) and row(s) of the 7709 connection matrix. Returns a list of channel numbers. Only one of the parameters ‘rows’ or ‘columns’ can be “all”

Parameters
  • rows – row number or list of numbers; can also be “all”

  • columns – column number or list of numbers; can also be “all”

  • slot – slot number (1 or 2) of the 7709 card to be used

check_errors()

Read all errors from the instrument.

Returns

List of error entries.

check_get_errors()

Check for errors after having gotten a property and log them.

Called if check_get_errors=True is set for that property.

If you override this method, you may choose to raise an Exception for certain errors.

Returns

List of error entries.

check_set_errors()

Check for errors after having set a property and log them.

Called if check_set_errors=True is set for that property.

If you override this method, you may choose to raise an Exception for certain errors.

Returns

List of error entries.

clear()

Clear the instrument status byte.

close_rows_to_columns(rows, columns, slot=None)

Closes (connects) the channels between column(s) and row(s) of the 7709 connection matrix. Only one of the parameters ‘rows’ or ‘columns’ can be “all”

Parameters
  • rows – row number or list of numbers; can also be “all”

  • columns – column number or list of numbers; can also be “all”

  • slot – slot number (1 or 2) of the 7709 card to be used

property closed_channels

Parameter that controls the opened and closed channels. All mentioned channels are closed, other channels will be opened.

property complete

Get the synchronization bit.

This property allows synchronization between a controller and a device. The Operation Complete query places an ASCII character 1 into the device’s Output Queue when all pending selected device operations have been finished.

config_buffer(points=64, delay=0)

Configures the measurement buffer for a number of points, to be taken with a specified delay.

Parameters
  • points – The number of points in the buffer.

  • delay – The delay time in seconds.

determine_valid_channels()

Determine what cards are installed into the Keithley 2700 and from that determine what channels are valid.

disable_buffer()

Disables the connection between measurements and the buffer, but does not abort the measurement process.

display_closed_channels()

Show the presently closed channels on the display of the Keithley 2700.

property display_text

A string property that controls the text shown on the display of the Keithley 2700. Text can be up to 12 ASCII characters and must be enabled to show.

get_state_of_channels(channels)

Get the open or closed state of the specified channels

Parameters

channels – a list of channel numbers, or single channel number

property id

Get the identification of the instrument.

is_buffer_full()

Returns True if the buffer is full of measurements.

property next_error

Get the next error in the queue. If you want to read and log all errors, use check_errors() instead.

open_all_channels()

Open all channels of the Keithley 2700.

property open_channels

A parameter that opens the specified list of channels. Can only be set.

open_rows_to_columns(rows, columns, slot=None)

Opens (disconnects) the channels between column(s) and row(s) of the 7709 connection matrix. Only one of the parameters ‘rows’ or ‘columns’ can be “all”

Parameters
  • rows – row number or list of numbers; can also be “all”

  • columns – column number or list of numbers; can also be “all”

  • slot – slot number (1 or 2) of the 7709 card to be used

property options

Property that lists the installed cards in the Keithley 2700. Returns a dict with the integer card numbers on the position.

read(**kwargs)

Read up to (excluding) read_termination or the whole read buffer.

read_binary_values(**kwargs)

Read binary values from the device.

read_bytes(count, **kwargs)

Read a certain number of bytes from the instrument.

Parameters
  • count (int) – Number of bytes to read. A value of -1 indicates to read the whole read buffer.

  • kwargs – Keyword arguments for the adapter.

Returns bytes

Bytes response of the instrument (including termination).

reset()

Resets the instrument and clears the queue.

reset_buffer()

Resets the buffer.

shutdown()

Brings the instrument to a safe and stable state

start_buffer()

Starts the buffer.

property status

Get the status byte and Master Summary Status bit.

stop_buffer()

Aborts the buffering measurement, by stopping the measurement arming and triggering sequence. If possible, a Selected Device Clear (SDC) is used.

property text_enabled

A boolean property that controls whether a text message can be shown on the display of the Keithley 2700.

triad(base_frequency, duration)

Sounds a musical triad using the system beep.

Parameters
  • base_frequency – A frequency in Hz between 65 Hz and 1.3 MHz

  • duration – A time in seconds between 0 and 7.9 seconds

wait_for(query_delay=0)

Wait for some time. Used by ‘ask’ to wait before reading.

Parameters

query_delay – Delay between writing and reading in seconds.

wait_for_buffer(should_stop=<function KeithleyBuffer.<lambda>>, timeout=60, interval=0.1)

Blocks the program, waiting for a full buffer. This function returns early if the should_stop function returns True or the timeout is reached before the buffer is full.

Parameters
  • should_stop – A function that returns True when this function should return early

  • timeout – A time in seconds after which this function should return early

  • interval – A time in seconds for how often to check if the buffer is full

write(command, **kwargs)

Write a string command to the instrument appending write_termination.

Parameters
  • command – command string to be sent to the instrument

  • kwargs – Keyword arguments for the adapter.

write_binary_values(command, values, *args, **kwargs)

Write binary values to the device.

Parameters
  • command – Command to send.

  • values – The values to transmit.

  • **kwargs (*args,) – Further arguments to hand to the Adapter.

write_bytes(content, **kwargs)

Write the bytes content to the instrument.