HP/Agilent/Keysight 34450A Digital Multimeter
- class pymeasure.instruments.agilent.Agilent34450A(adapter, name='HP/Agilent/Keysight 34450A Multimeter', **kwargs)
Bases:
SCPIUnknownMixin
,Instrument
Represent the HP/Agilent/Keysight 34450A and related multimeters.
dmm = Agilent34450A("USB0::...") dmm.reset() dmm.configure_voltage() print(dmm.voltage) dmm.shutdown()
- beep()
Sounds a system beep.
- property capacitance_auto_range
Control auto ranging for capacitance.
- property capacitance_range
Control the capacitance range in Farads, which can take values 1E-9, 10E-9, 100E-9, 1E-6, 10E-6, 100E-6, 1E-3, 10E-3, as well as “MIN”, “MAX”, or “DEF” (1E-6). Auto-range is disabled when this property is set.
- configure_capacitance(capacitance_range='AUTO')
Configures the instrument to measure capacitance.
- Parameters:
capacitance_range – A capacitance in Farads to set the capacitance range, can be 1E-9, 10E-9, 100E-9, 1E-6, 10E-6, 100E-6, 1E-3, 10E-3, as well as “MIN”, “MAX”, “DEF” (1E-6), or “AUTO”.
- configure_continuity()
Configures the instrument to measure continuity.
- configure_current(current_range='AUTO', ac=False, resolution='DEF')
Configures the instrument to measure current.
- Parameters:
current_range – A current in Amps to set the current range. DC values can be 100E-6, 1E-3, 10E-3, 100E-3, 1, 10, as well as “MIN”, “MAX”, “DEF” (100 mA), or “AUTO”. AC values can be 10E-3, 100E-3, 1, 10, as well as “MIN”, “MAX”, “DEF” (100 mA), or “AUTO”.
ac – False for DC current, and True for AC current
resolution – Desired resolution, can be 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- configure_diode()
Configures the instrument to measure diode voltage.
- configure_frequency(measured_from='voltage_ac', measured_from_range='AUTO', aperture='DEF')
Configures the instrument to measure frequency.
- Parameters:
measured_from – “voltage_ac” or “current_ac”
measured_from_range – range of measured_from. AC voltage can have ranges 100E-3, 1, 10, 100, 750, as well as “MIN”, “MAX”, “DEF” (10 V), or “AUTO”. AC current can have ranges 10E-3, 100E-3, 1, 10, as well as “MIN”, “MAX”, “DEF” (100 mA), or “AUTO”.
aperture – Aperture time in Seconds, can be 100 ms, 1 s, as well as “MIN”, “MAX”, or “DEF” (1 s).
- configure_resistance(resistance_range='AUTO', wires=2, resolution='DEF')
Configures the instrument to measure resistance.
- Parameters:
resistance_range – A resistance in Ohms to set the resistance range, can be 100, 1E3, 10E3, 100E3, 1E6, 10E6, 100E6, as well as “MIN”, “MAX”, “DEF” (1E3), or “AUTO”.
wires – Number of wires used for measurement, can be 2 or 4.
resolution – Desired resolution, can be 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- configure_temperature()
Configures the instrument to measure temperature.
- configure_voltage(voltage_range='AUTO', ac=False, resolution='DEF')
Configures the instrument to measure voltage.
- Parameters:
voltage_range – A voltage in Volts to set the voltage range. DC values can be 100E-3, 1, 10, 100, 1000, as well as “MIN”, “MAX”, “DEF” (10 V), or “AUTO”. AC values can be 100E-3, 1, 10, 100, 750, as well as “MIN”, “MAX”, “DEF” (10 V), or “AUTO”.
ac – False for DC voltage, True for AC voltage
resolution – Desired resolution, can be 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- property current_ac_auto_range
Control auto ranging for AC current.
- property current_ac_range
Control the AC current range in Amps, which can take values 10E-3, 100E-3, 1, 10, as well as “MIN”, “MAX”, or “DEF” (100 mA). Auto-range is disabled when this property is set.
- property current_ac_resolution
Control the resolution in the AC current readings, which can take values 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- property current_auto_range
Control auto ranging for DC current.
- property current_range
Control the DC current range in Amps, which can take values 100E-6, 1E-3, 10E-3, 100E-3, 1, 10, as well as “MIN”, “MAX”, or “DEF” (100 mA). Auto-range is disabled when this property is set.
- property current_resolution
Control the resolution in the DC current readings, which can take values 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, and “DEF” (3.00E-5).
- property frequency_aperture
Control the frequency aperture in seconds, which sets the integration period and measurement speed. Takes values 100 ms, 1 s, as well as “MIN”, “MAX”, or “DEF” (1 s).
- property frequency_current_auto_range
Control whether auto ranging for AC current in frequency measurements is enabled.
- property frequency_current_range
Control the current range in Amps for frequency on AC current measurements, which can take values 10E-3, 100E-3, 1, 10, as well as “MIN”, “MAX”, or “DEF” (100 mA). Auto-range is disabled when this property is set.
- property frequency_voltage_auto_range
Control whether auto ranging for AC voltage in frequency measurements is enabled.
- property frequency_voltage_range
Control the voltage range in Volts for frequency on AC voltage measurements, which can take values 100E-3, 1, 10, 100, 750, as well as “MIN”, “MAX”, or “DEF” (10 V). Auto-range is disabled when this property is set.
- property mode
Control the measurement mode of the multimeter. Can be “current”, “ac current”, “voltage”, “ac voltage”, “resistance”, “4w resistance”, “current frequency”, “voltage frequency”, “continuity”, “diode”, “temperature”, or “capacitance”.
- property resistance
Get a resistance measurement in Ohms for 2-wire configuration, based on the active
mode
.
- property resistance_4w
Get a resistance measurement in Ohms for 4-wire configuration, based on the active
mode
.
- property resistance_4w_auto_range
Control auto ranging for 4-wire resistance.
- property resistance_4w_range
Control the 4-wire resistance range in Ohms, which can take values 100, 1E3, 10E3, 100E3, 1E6, 10E6, 100E6, as well as “MIN”, “MAX”, or “DEF” (1E3). Auto-range is disabled when this property is set.
- property resistance_4w_resolution
Control the resolution in the 4-wire resistance readings, which can take values 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- property resistance_auto_range
Control auto ranging for 2-wire resistance.
- property resistance_range
Control the 2-wire resistance range in Ohms, which can take values 100, 1E3, 10E3, 100E3, 1E6, 10E6, 100E6, as well as “MIN”, “MAX”, or “DEF” (1E3). Auto-range is disabled when this property is set.
- property resistance_resolution
Control the resolution in the 2-wire resistance readings, which can take values 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- property voltage_ac_auto_range
Control auto ranging for AC voltage.
- property voltage_ac_range
Control the AC voltage range in Volts, which can take values 100E-3, 1, 10, 100, 750, as well as “MIN”, “MAX”, or “DEF” (10 V). Auto-range is disabled when this property is set.
- property voltage_ac_resolution
Control the resolution in the AC voltage readings, which can take values 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).
- property voltage_auto_range
Control auto ranging for DC voltage.
- property voltage_range
Control the DC voltage range in Volts, which can take values 100E-3, 1, 10, 100, 1000, as well as “MIN”, “MAX”, or “DEF” (10 V). Auto-range is disabled when this property is set.
- property voltage_resolution
Control the resolution in the DC voltage readings, which can take values 3.00E-5, 2.00E-5, 1.50E-6 (5 1/2 digits), as well as “MIN”, “MAX”, or “DEF” (1.50E-6).